3775 Alcohol-Based Condensation NucleiParticle Counteris a general-purpose counter that detects suspended airborne particles as small as 4nm. It guarantees highly accurate detection over a wide concentration range of 0-107 particles/cm3. This makes the CPC powerful and suitable for a wide range of applications.
3775 is the successor to the Model 3022A, which features continuous real-time calibrated extended single particle counting at concentrations up to 50,000 particles/cm3. Higher concentrations are required for photometer mode, and the Model 3775 also features faster conversion between the two counting modes. It features a color VGA display, spill-proof design with water removal, data logging and storage via removable memory card, and USB and Ethernet ports. It includes a built-in SMPS? compatible system and our Aerosol Instrument Manager? software.
Lower particle size detection limit of 4nm
Fast response to rapid changes in aerosol concentration (T95 @ 4 seconds)
Extended single particle counting up to 5 x 104 particles/cm3 with continuous real-time coincidence correction
Smooth transition to photometric mode counting, which counts up to 107 particles/cm3
Butanol-friendly features, including anti-overflow design, water removal system, butanol odor absorber, and improved resistance to optical immersion after instrument tilt or input flow restriction
Built-in data logging and data storage using removable memory card
USB and Ethernet ports available
Built-in SMPS compatible system (now standard!)
Automatic recovery from power outages
Detachable saturator core for easy transport and maintenance
Particle concentration, concentration-time comparison, total count, or instrument status displayed on front panel color LCD display