The TSI 9110 Laser Particle Counter can accurately measure particles down to 0.100 microns. It uses He-Ne laser technology that greatly improves the signal-to-noise ratio. It is fully compliant with ISO-14644-1 and is the best choice for verifying Class I and Class II clean rooms. This instrument can be used for clean room monitoring, process monitoring, and filter membrane testing.
The TSI 9110 Particle Counter can generate ISO14644-1 pass or fail reports. It can be used not only as a stand-alone particle counter, but also compatible with online particle monitoring systems such as FMS. It is calibrated using NIST-traceable PSL microspheres, recognized particle inspection standards, TSI advanced sifters, and condensation particle counters.
Sample volume 28.3L/min (1.0cfm)
Accurate detection of particles as small as 0.100 μm
Simultaneous detection of up to eight channels of particle data
Meets all requirements of ISO 21501-4
Best choice for Class I and Class II cleanroom testing
Fully compliant with ISO 21501-4
Generates compliance reports for ISO-14644-1, EU GMP Annex1 and FS209E
External certification reports output via TRAKPRO Lite Secure software
Enhanced HeNe laser
Highly repeatable and reliable testing/data
Data storage:250 zones, 999 locations and 10,000 samples
USB, Ethernet and pulse outputs
Industrial grade high decibel alarm
Allows storage and playback of specific tests/configurations
Displays up to three environmental parameters
Seamlessly integrates with M32-01 multi-point sampler
Cleanroom classification up to ISO I class
Model | TSI-9110 |
Particle size range | 0.1-10μm |
Particle size channel | 01:0.1,0.15,0.2,0.25,0.3,0.5,1.0,5.0μm |
Particle size resolution | 01:at 0.2μm<15%> |
Counting efficiency | 01:50% when 0.1μm, 100% when greater than 0.15μm (in compliance with ISO21501-4 and JIS) |
Concentration limit | 100,000 pieces/ft3 (3,500,00 pieces/m3), 10% overlap loss |
Laser source | Enhanced HeNe Laser |
Zero point | 1 per 5 minutes (compliant with ISO21501-4 and JISB9921) |
Sampling volume | 28.3L/min (1.0 cfm) |
Flow control | Electronic automatic closed loop (patented flow control technology) |
External pipeline | External pipeline up to 8 meters |
Sampling mode | Manual, automatic, buzzer. Cumulative/Difference;Count or Concentration |
Sampling Time | 1 second to 99 hours |
Sampling Frequency | 1 to 9999 cycles or continuous type |
Exhaust | Built-in HEPA filter |
Power source | Built-in pump |
Communication mode | Communication mode Modbus®RTU via Ethernet or USB interface |
Data storage capacity | 10,000 data records, including date, time, six channels, flow, sample ID, sampling flow |
Sound warning | Built-in, greater than 85dB at 1 meter (adjustable) |
Status indication | Low battery, flow, laser |
Alarm output | Touch point, closed when alarm |
Alarm limit | Different channels are programmable |
Display | VGA 14.5-cm touch screen |
Language | Simplified Chinese, English, Japanese, German, French, Spanish and Italian |
Software | Compatible with TrakProTM Lite and FMS5 software |
Flow control | Electronic automatic closed loop (patented flow control technology) |
External pipeline | External pipeline up to 8 meters |
Sampling mode | Manual, automatic, buzzer. Cumulative/Difference;Count or Concentration |
Sampling Time | 1 second to 99 hours |
Sampling Frequency | 1 to 9999 cycles or continuous type |
Exhaust | Built-in HEPA filter |
Power source | Built-in pump |
Communication mode | Communication mode Modbus®RTU via Ethernet or USB interface |
Data storage capacity | 10,000 data records, including date, time, six channels, flow, sample ID, sampling flow |
Sound warning | Built-in, greater than 85dB at 1 meter (adjustable) |
Status indication | Low battery, flow, laser |
Alarm output | Touch point, closed when alarm |
Alarm limit | Different channels are programmable |
Display | VGA 14.5-cm touch screen |
Language | Simplified Chinese, English, Japanese, German, French, Spanish and Italian |
Software | Compatible with TrakProTM Lite and FMS5 software |
Storage range | 0-50℃, 98% non-condensing |
Standard | ISO 21501-4, CE, JIS B9921 |
Accessories included | Power cord, battery, injection port, isokinetic probe, tripod, pipeline, filter membrane, printing paper, USB data cable |
Optional accessories | Filter scanning probe, TSI wind probe, temperature and humidity probe, spare battery, dual-port external charger, constant speed probe |
Particle test in clean workshop
Air particle research
Exposure assessment
Indoor air quality assessment
Filter performance test
Cleanliness evaluation and pollutant migration research, etc.