The TSI 9500 Laser Dust Particle Counter offers best-in-class features and functionality. For ISO14644-1 and EU GMP cleanliness class certification required for a 1 cubic meter air sample takes just 10 minutes! Particle counts can be easily displayed on the display, downloaded using TRAKPRO Lite Secure software, or printed directly using the instrument's integrated printer.The Model 9500 counter is compatible with TSI's precision ventilation probes that measure air speed/temperature/humidity, providing a complete solution for your certification. Used in conjunction with one of the optional sampling hoses up to 0.6 meters (26 inches) in length and an electronic filter scanning probe, it provides a better instrument for cleanroom HEPA filter testing.
TSI 9500 Laser Dust Particle CounterThe TSI 9500 Laser Dust Particle Counter generates ISO14644-1, EU EUGMP Annex 1, and FS2O9E compliant, downloadable or printable pass/fail reports via the integrated printer and TRAKPRO Lite Secure software. The particle counter's versatile configuration allows for easy storage and uploading of data as needed. The product can be used as a stand-alone particle counter or integrated into an online plant monitoring system like TSI's FMS 5. Strict adherence to compliance with ISO 21501-4 standards. Calibration is performed using NIST-traceable PSL microspheres, recognized particle inspection standards, TSI's advanced sifter and condensation particle counters.
Sampling volume 100L/min (3.53cfm)
Measures six channels of data
Fully compliant with ISO 21501-4. span>
Data storage:250 zones, 999 locations and 10,000 samples
Data transfer over USB and Ethernet
Industrial audible and visual alarm
High-definition touch screen, easy-to-use graphic interface, Simplified Chinese
Generates compliance reports for ISO-14644-1, EU GMP Annex1 and FS209E
Through the External certification report output by TRAKPRO Lite Secure software
Allows storage and playback of specific tests
>Displays up to 3 environmental parameters
Stainless steel housing
Light weight
Stand-alone or integrated into an online monitoring system
™ lt;td style="word-break:break-all;"> 02:at 0.5μm<15%>Exhaust <
Model | TSI-9500 | |||
Particle Size Range | 0.5-25μm | |||
Particle size channel | 02:0.5,0.7,1.0,3.0,5.0,10.0 μm (additional channels optional) | |||
Particle Size Resolution | ||||
Counting efficiency | 02:0.5μm 50%, greater than 0.5μm. 50% at 0.5μm, 100% at >0.75μm (in accordance with ISO21501-4 and JIS) | |||
Concentration Limit | 5% loss at >400,000 pcs/ft3 | |||
Laser Source | Long-life Laser Diode | |||
Zero Point | Every 5 minutes<1> | |||
Sampling volume | 100L/min(3.53cfm) | |||
Flow Control | Electronic Auto-Close Loop (Patented Flow Control Technology) | |||
External Pipeline | External piping up to 8 meters | |||
Sampling Mode | Manual, Auto, Beep. Totalization/Difference;Count or Concentration | |||
Sampling Time | 1 sec to 99 hrs | |||
Sampling Frequency | 1 to 9999 Cyclic or Continuous | |||
Exhaust | Sampling Frequency | Built-in HEPA Filter Membrane | ||
Power Source | Power Source | Power Source | Built-in pump | |
Built-in pump | ||||
Communication mode | Communication mode Modbus®RTU via Ethernet or USB Interface | |||
Data storage | 10,000 data records including date, time, six channels, flow rate, sample ID, sample flow rate, transfer data via USB or use the TraxProTM software download | |||
Audible Warnings | 内置,1米时大于85dB(可调整) | |||
Low Battery, Flow, Laser | ||||
Alarm Output | Sensor Point, Off On Alarm | |||
Alarm Limits Programmable for different channels | ||||
Alarm Limits | Alarm LimitsDisplay | VGA | VGA | |
Language | Simplified Chinese, English, Japanese, German, French, Spanish, and Italian | |||
软件 | Compatible with TrakProTM Lite and FMS5 software | |||
Printer | Built-in thermal printer (no printer optional) | |||
Printout | Selectable for all languages | |||
Environmental Sensor Interface | Support TSI Probe Models:960,962,964,966 (Air Flow/Temperature/Humidity) and Temperature-Humidity Probes | |||
Unit No. | Configurable IP address | |||
Security | Secondary password protection | |||
Local ID | to 999 positions;each 16 characters long | |||
Report< | Provides a comprehensive range of products and services based on ISO14644-1, EU GMP and  amp;nbsp;FS209E pass/fail reports | |||
Calibration | Calibration | Traceable to NIST's TSI Calibration System | ||
Calibration Frequency | Recommended at least annually | |||
Volume (Hx Lx W) | 24.1cmx23cmx23.2cm(excluding isovelocity sampling port) | |||
Weight | 6.44 kg (two batteries) | |||
Power | 110 to 240VAC, normal power supply | |||
Battery | Removable rechargeable lithium battery, 2 battery slots | |||
Battery life | Can be used continuously for more than 3 hours | |||
Charging time | 4 hours (when idle) | |||
Surface Material | Stainless Steel | |||
Operating range | 2-35°C, 20%-95% non-condensing | |||
0-50℃, 98% non-condensing | ||||
Standard | Standard | Standard | ISO 21501-4, CE, JIS B9921 | |
Included Accessories | Quick Start Manual, CD-ROM Operator's Manual, Power Cord, Battery, Inlet, Isokinetic Probe, Tripod, Tubing, Filter Membrane, Printout, USB Cable and TrakProTM Lite Software | |||
Optional Accessories Electronic Filtration Scanning Probe, Basic Filtration Scanning Probe, TSI Windage Probe, Temperature and Humidity Probe, Spare Battery, Dual Port External Charger, Isobaric Probe, Isobaric Sampling Inlet, Sampling Tubes, Packaging Box |
Particle testing in clean rooms
Airborne particles Research
Leakage Assessment
Leakage Assessment
Indoor Air Quality Assessment
Filter Performance Test
Filter Performance Test
Filter Performance Test
Cleanliness Evaluation and Pollutant Migration Studies
WeChat Account