U.S. hash insoluble particles detector HIAC9703+
-Simplifies the sampling process, using fewer samples, reducing contamination and sampling costs
-Supports 21CFR Part II and Pharma Spec 2.0
-Meets new USP<789> testing for ophthalmic products.
-Easier monitoring. Industry's smallest pre-sampler
-Small tube clamp ensures that samples do not spill during testing.
- Automatically performs routine rinses and cleanses, enjoying more time away from the test.
Performance:
-Measurement of particle size (um) : 0.1-350um , HRLD
5 different measuring range probes to choose from - and Micro Count
-Sample viscosity < 20 cp, temperature range 5-40°C
-Sample flow rate 10-100ml/min
-Flow rate accuracy >95%, HRLD accuracy >95
Customized routine testing and reporting
-Particle size, sampling volume and running test methods can be customized to meet your unique application requirements.
-User configurable reports
-Enhanced flexibility to configure security access
-Multiple security access levels can be administratively defined
-Meets the requirements of 21 CFR Part 11 regulations.
Injectable Drugs
-USP <788
-EP <2. 9. 19
-JP<11>, <24
-KP<52
Ophthalmic Drugs & nbsp;   nbsp; & & nbsp;
-USP<789>
Intake medication
-USP<788>/Levosalbutamol ingestion drug
Manage all your application needs
Designed to handle from very small volumes (1.0ml) to very large volumes ( 1,00ml ) to very large volumes (1,00ml)
-Injectable small doses
-Injectable large doses
-Development
-Trouble-shielded
-Stability testing
Easy configuration for maximum flexibility
-Interchangeable sensors from 0.54um to 600um
-Interchangeable sampling probes
-Syringes are interchangeable - depending on the sampling star.
-Flow rates from 10 ml/min to 100 ml/min can be configured.
Elimination of instabilities based on data anomalies
-Alarm notification when air bubbles may affect counting results
-Referral to the operator if the sensor is contaminated
-Advance notification when routine service or calibration is required
-Operator notification when routine upgrades are available
-All alarms provide recommended actions to be taken
Eliminates your fear of data loss
-Recovery of lost data protects test data in the event of an unplanned power failure.
-PharmSpec allows operators to schedule data backups to a secure network site.
-Telephone notification when your instrument requires routine maintenance
Instrument Validation Services
-SOF Help and IQ/OQ Validation Services
-Apply our expertise to make it easier and simpler for you to use.
-Ensure the accuracy of your instrument with archived validations and IS021051-3 regulation calibrations.
9703+ Sampler
Operating temperature range 5-40°C (40-104°F), 0-80% relative humidity, non-condensing
Sample temperature range 5-40°C(40-104°F)
Sample viscosity <50 cP
Power Requirements 100/230VAC, 50/60 Hz
Maximum power 67W for 115/230VAC
Dimensions 343WX337 DX482 H mm (13.5X 13.3 X 19.0 inches)
Weight 10.7kg (23.6 lbs)
Sampling Vial Clearance 153 mm (6.02 in. j
Volume Accuracy >95%
Flow Rate Accuracy >95%
Sample flow rate 10-100 ml/min: actual flow rate of the system is determined by the flow rate of the sensor
Probe pre-draw volume 0.091 ml for 81 LX 1.2IDmm (3.2X0.047 inches)
0.172 ml (6.1X0.047 inches) for a 154 LX 1.2IDmm probe
1.57 ml (5.5X0.25 inches) for 140 LX 6.35IDmm probe - large bore
PharmSpec 3 Software
Processor Pentium1.0GHz
Operating System Windows XP, Service Pack 2 or higher.
Windows Vista, Service Pack 2 or higher, Windows 7
Memory 512MB or RAM
Hard disk 2GB of space available on the hard disk
Optical drive CD-ROM drive, DVD-ROM drive, USB port or network connection for software installation
Display High-quality VGA video adapter and monitor, 1024x768 or higher resolution
Power supply Continuous power supply (UPS)
Removable CD-ROM drive Data can be backed up, e.g., recordable CD-ROM or network phone, can back up computer hard disk data
Printer nbsp;Color printer with 300 dpi resolution.
COM port USB port for connection to particle counting sensors.
Software Internet Explorer 6.0 and Microsoft Excel, Word or Adobe Acroabt (with access to PharSpec reports).
Typical System Configurations and Other Options Available
Item Number Description: 2089105-01 9703
2089105-01 9703+ Sampler
2089250-150 HRLD-150 Sensor
700060GP PharmSpec 3 Software
DOC026.53.8009 Validation File-English (other languages also available)
2089007-025 25ml Calibration Flow Rate (other flow rates also available)
Standard Sensors Ranges
HRLD-150 1.3 um -150 um
HRLD-150JA 1.3 um -150 um (chemically compatible)
HRLD-400 2.0 um -400 um
HRLD-600JS 2.0 um -600 um (chemical compatibility)

WeChat Account

WeChat Video Account