American Hach Insoluble Particle Detector HIAC9703+
•Simplify the sampling process, use less samples, reduce pollution and sampling costs
•Support 21CFR Part II and Pharma Spec 2.0
•Meet the new USP<789<>> test for ophthalmic products
•Easier monitoring. The smallest pre-extraction volume in the industry
•Small test tube clamp ensures that the sample will not splash during testing
•Automatically perform routine flushing and cleaning, enjoy more time away
Performance:
•Measurement of particle size (um): 0.1-350um, HRLD
• and Micro Count 5 different measuring range probes to choose from
• Sample viscosity < 20 cp, temperature range 5-40°C
• Sampling flow rate 10-100ml/min
•Flow accuracy>95%, volume accuracy>95%
Customize routine tests and reports
•Customize particle size, sampling volume and run test methods according to the user's unique application requirements
•Users can configure reports
•Enhanced flexibility can configure security access
•Can manage and define multiple security access levels
•Meet the requirements of 21 CFR Part 11 regulations
Injectable drugs
•USP<788>
•EP<2. 19="" 9.="">
•JP<11>,<24>
•KP<52>
Eye drops ;
•USP<789>
Intake medication
•USP<788>/Levosalbutamol intake medication
Manage all your application needs
Designed to handle the full range from very small volumes (1.0ml)to very large volumes (1,00ml)
•Injectable small dose star
•Injectable large dose
•Research and development
•Troubleshooting
•Stability testing
Easy configuration for maximum flexibility
•Interchangeable sensors from 0. 54um to 600um
•Interchangeable sampling probes
•Interchangeable syringes - depending on the sampling star
•Flow rate can be configured from 10ml/min to 100ml/min
Eliminate instability based on data anomalies
•Alarm notification will be issued when bubbles may affect the counting results
•If the sensor is contaminated, the operator can be reminded
•When routine service or calibration is required, advance notice will be given
•When routine upgrades are available, operators will be notified
•All alarms can provide recommended actions
Eliminate your worries about data loss
•The lost data recovery function can protect test data in the event of an unexpected power outage
•PharmSpec allows operators to schedule data backup to a secure network site
•When your instrument requires routine maintenance, you will be notified by phone
Instrument Validation Service
•SOF assistance and IQ/OQ validation service
•Apply our expertise to make your use easier
•The accuracy of the instrument can be ensured by using archived verification and IS021051-3 regulatory calibration
9703+ sampler
Operating temperature range 5-40°C(40-104°F), relative humidity 0-80%, non-condensing
Sample temperature range 5-40°C(40-104°F)
Sample viscosity <50 cp
Power Requirements Hz
Maximum power 115/230VAC is 67W
Size H mm (13.5X 13.3X 19.0 inches)
Gap of sampling bottle 153mm(6.02 inchesj
Volume accuracy >95%
Flow rate accuracy >95%
Sample flow rate 10-100ml/min: The actual flow rate of the system is determined by the flow rate of the sensor
The pre-pump volume of the probe is 0.091 ml for the 81LX1.2IDmm probe (3.2X0.047 inches)
0.172 ml for the 154 LX 1.2IDmm probe (6.1X0.047 inches)
1.57 for the 140LX6.35IDmm probe ml (5.5X0.25 inches)-large hole
PharmSpec 3 Software
Processor ; Pentium1.0GHz
Operating system Windows XP, Service Pack 2 or later,
Windows Vista, Service Pack 2 or later, Windows 7
Memory 512MB or RAM
Hard disk The hard disk can provide 2GB of space
Optical drive CD-ROM, DVD-ROM, USB port or network connection, for installing software
Display Good quality VGA video adapter and monitor, 1024X768 or higher resolution
Power supply Continuous power supply (UPS)
Removable optical drive can back up data, such as recordable CDs or network servers, can back up computer hard disk data
COM port USB port can be connected to the particle counting sensor
Software Internet Explorer 6.0 and Microsoft Excel, Word or Adobe Acroabt (to access PharSpec reports)
Typical System Configuration - Other Options Available
Article number Description
2089105-01 9703+sampler
700060GP PharmSpec 3 Software
DOC026.53.8009 Validation Document - English (Other languages available)
2089007-025 25ml calibration flow rate (other flow rates are also available)
Standard sensor Range
HRLD-150 1.3 um -150 um
HRLD-150JA 1.3 um -150 um (chemical compatibility)
HRLD-400 2.0 um -400 um
HRLD-600JS 2.0 um-600um (chemical compatibility)
Regulatory Certification
CE certified according to EMC and Low Voltage Directives: Class I laser product, compliant with IEC/E N 60825-1 and 21CFR1040.10 Laser noise is 50** (**Contact the manufacturer for complete regulatory details)
WeChat Account
WeChat Video Account