Applicable to environmental monitoring stations, third-party testing institutions, scientific research institutions, semiconductors, new energy, biomedicine and related enterprises for the pre-concentration treatment system of volatile organic compounds in gas samples.
Technical parameters
Detection limit: 0.1ppbv
Injection range: 4~2000ml
Working environment: Temperature 0~50℃; Humidity 0-95%RH
Concentration rate: >1000:1
Ultra-low temperature control accuracy: ±2℃
Power supply voltage: 220V/50Hz±10%
Reproducibility: ≤3%RSD
Maximum heating rate: 10000℃/min
Dimensions: width 300x height 450x depth 450mm
Performance characteristics
1 The three-stage cold trap method has strong applicability
The fast sample transfer mode is adopted to enable the VOCs in the focusing trap to quickly enter the chromatographic column, and achieve secondary focusing during transmission in the column tube, so as to ensure a sharper peak shape. It complies with the current national standard HJ759-2023, for 65 VOCs in various scene samples such as ambient air, indoor air, and exhaust gas from pollution sources.
2 Wide applicability of high and low concentrations
The system has a built-in quantitative loop, with a concentration rate of more than 1,000 times, which meets the injection application requirements of high-concentration sample analysis, effectively improves the detection limit of analytical instruments, and has an accuracy better than the national standard.
3 Fully automated injection
Sequence editing, automatic system operation, and automatic error reporting. When the number of samples is higher than 19, 4 automatic samplers can be connected to achieve unattended injection. Simple operation, record process data, and report printing to support quality control.
4 Good compatibility and powerful functions
It can be paired with imported or domestic sampling cans and air bags, and is compatible with mainstream GC or GC-MS on the market; it can independently inject 4 channels, and can connect multiple groups of automatic samplers; it can establish a unique analysis method based on the characteristics of the compound in the preset analysis method.
5 Stable operation and intelligent baking
The composition of the gas path is richer, more flexible, and more variable, and it can adapt to more advanced analysis methods. And it is more accurate and reasonable in gas path switching and time control. Rich gas path changes and precise control can greatly improve the accuracy of the instrument. When the GC delay time reaches 5 minutes before the end, preheat the cold trap 3 in advance to remove residual moisture to ensure the consistency of each baking, avoid the influence of impurities and water on the system and analysis results, and improve the stability of the analysis.
6. Low analysis cost
The new electronic refrigeration and cooling module technology has gotten rid of the traditional liquid nitrogen cooling and VOCs capture technology. It simplifies the analysis workflow, saves consumables and improves work efficiency.