Particle counter calibration by the Metrology Institute vs. factory calibration
- 2023-02-08
- 1417
- Esky Purify
Particle counter metrology school calibration vs. factory calibration
Serial number | Calibration Process | Service Content | Original Calibration | VS | Metrology institute or third party testing organization |
1 | Functional check | Check the health condition of the machine, whether it is functioning normally or not. | The machine is checked to be in good condition and calibration is performed. Check the machine for faults and issue a fault detection report. | VS | Faulty machine, return calibration instrument. VS Faulty machine, return calibration instrument. VS Faulty machine, perform comparison calibration test (portable). On-line sensors need to be read by monitoring software, which is not available at the institute, so the test cannot be read directly and a comparison calibration cannot be performed. |
2 | Status check before calibration | 1. Warm up;2. Check the Calibration Voltage status;3. Check the flow;4. Measure the noise;5. Measure the gain. | Clean the sensor to restore the Calibration Voltage to the normal value, confirm the parameters are correct and start the calibration work. | VS | Sensor Calibration Voltage exceeds the standard, the sensor alarm, can not be compared to the calibration, the calibration instrument will be returned (portable) can not complete other test items. |
3 | Internal Inspection | 1. Laser performance check;2. Warm up;3. Confirm flow rate; |
Calibration standard procedure | VS | Not available for metrology school calibration |
4 | Calibration using a single standard microparticle | Engineering plastics such as ABS/PVC | Calibration standard program | VS | Measurement Institute to hit the standard particles, do not specifically specify how large particle size particles to hit;standard machine and to be tested machine for comparison test (± 30%). |
5 | Other calibration program check | 101Pa@2.5m/s | Calibration standard program | VS | No such item for metrology school calibration |
6 | Machine operation test | -20℃~51℃ | Calibration Standard Program | VS | No such content in the calibration of measuring institutions |
7 | Reporting information | Issuance of original calibration certificate | Calibration standard program | VS | Issuance of Metrology School Correction Report |
8 | Man-hour | Labor time to perform 1 sensor | 2.5~8 hours | VS | 0.5 hours |